The doseBadge Noise Dosimeter is a high performance noise measurement instrument that meets the standards required by the Noise at Work Regulations as well as other International occupational noise regulations.
The specifications and performance information for the doseBadge are shown below.
Applicable Standards
CR:110A doseBadge
IEC 61252:1993 Personal Sound Exposure Meters
ANSI S1.25:1991 Personal Noise Dosemeters Class Designation 2AS-90/80-5
RC:110A Reader Unit
Internal Acoustic Calibrator to IEC 60942:2003 Class 2
Measurement Range
(Typical) 70dB(A) to 130dB(A) RMS
120dB(C) to 140dB(C) Peak
Measurement Functions
The CR:110A doseBadge and RC:110A Reader Unit support Dual Channel measurements.
The information provided by the doseBadge consists of the Overall Measurement Data, Channel 1 measurement data and Channel 2 measurement data.
If the doseBadge is configured with Channel 1 and Channel 2 having identical parameters, the dBLink3 software will only display Channel 1.
Overall Measurement Data
doseBadge Configuration (Badge Serial Number, Date & Time)
Calibration Record
Measurement Duration
Highest Peak(C) Sound Level during the measurement
Overload Exceedence
115dB(A) Maximum Sound Level Exceedence Battery Status
Channel 1 Measurement Data
For 3dB Exchange Rate:
LAeq, LEX,8h, LAE, % Dose, Exposure (Pa2h) Estimated % Dose, Estimated Exposure (Pa2h)
For 4dB & 5dB Exchange Rates (and configurations with Time Weighting or Threshold): LAVG, TWA, % Dose, Estimated % Dose
1 Minute Time History of:
LAeq (3dB) or LAVG (4dB or 5dB)
Peak(C) Level
Battery Level
Channel 2 Measurement Data
LAeq, LEX,8h, LAE, % Dose, Exposure (Pa2h)
Estimated % Dose, Estimated Exposure (Pa2h)
1 Minute Time History of LAeq (3dB) & Peak(C)
Weightings Frequency
‘A’ for all RMS measurements ‘C’ for Peak Sound Pressure
Configuration Options
Channel 1:
Independent User Configuration of:
Exchange Rate: 3dB, 4dB or 5dB
Criterion Level: 80dB, 85dB, 90dB
Criterion Time: 8hrs, 12hrs, 16hrs, 18hrs
Threshold: None, 80dB, 90dB
Time Weighting: None, ‘S’ (Slow)
Channel 2 Preset to:
Exchange Rate:3dB
Criterion Level: 85dB
Criterion Time: 8hrs
Threshold: None
Time Weighting: None
Memory
CR:110A doseBadge
The CR:110A doseBadge can store up to 24 hours of data in a single measurement
RC:110A Reader
Up to 999 Individual doseBadge Measurements
Power
RC:110A doseBadge
Internal NiMH Battery. Typical Battery Life 30 hours @ 80dB
RC:110A Reader 2 x AA/LR6 with Auto Power Switch Off
CU:101A Charger Unit
CU:195A Mains Power Supply. Fast Charge Option
Output
CR:110A doseBadge
Wireless Infrared to RC:110A Reader Unit
RC:110A Reader
USB 2.0 (which also provides power to the RC:110A Reader)
Dimensions
CR:110A doseBadge
Microphone Apex Ø13.0mm, Base Ø47mm, Height 38mm
Weight
CR:110A doseBadge 51g (1.8oz)
RC:110A Reader 400g (14oz)
Environmental
Temperature
-10°C to +50°C Operating
-20°C to +60°C Storage
Humidity Up to 95% RH Non-Condensing
Software
dBLink3 supplied as standard with license free installation and free of charge upgrades available from the Cirrus website
Intrinsic Safety Certification
The CR:110AIS doseBadge is available with Intrinsic Safety Certification for both Mining and Non-Mining Applications. The CR:110AIS doseBadge carries ATEX, IECEx, EEx, & FM approvals as described below.
| Group I Non-Mining Applications | Group II Mining Applications |
|
|---|---|---|
| ATEX | CE Ex II 1 G | CE Ex I M1 |
| IECEx | IECEx BAS 06.0084 Ex ia IIC T4 (-20°C ≤ Ta ≤ +60°C) | IECEx BAS 06.0083X Ex ia I (-20°C ≤ Ta ≤ +60°C) |
| EEx | EEx ia IIC T4 (-20°C ≤ Ta ≤ +60°C) | EEx ia I (-20°C ≤ Ta ≤ +60°C) |
| FM | IS / I / 1 / ABCD / T4 Ta = 60°C; Type IP54, I / 0 / AEx ia / IIC / T4 Ta = 60°C; Type IP54 |